| Sample Volume |
0.5uL to 2uL, recommended 2uL |
| Wavelength Range |
180nm to 910nm |
| Wavelength Accuracy |
±1nm |
| Wavelength Resolution |
≤1.5nm, FWHM at Hg 253.7nm |
| Absorbance Range |
0.04 to 750, at 260nm, equivalent to 10mm optical path |
| Absorbance Precision |
0.002Abs, 1mm pathlength |
| Absorbance Accuracy |
±1%, 7.332 at 260nm |
| Detection Time |
less than 6 seconds |
| Pathlengths |
0.02mm, 0.03mm, 0.1mm, 0.2mm, 1mm |
| Light Source |
Pulsed xenon lamp |
| Light Lifetime |
more than 1 billion flashes |
| Detection Concentration Range |
2 to 37,500ng per uL dsDNA
1.32 to 24,750ng per uL ssDNA
1.6 to 30,000ng per uL RNA |
| Protein Detection Range |
BSA: 0.06 to 1119mg per mL
IgG: 0.03 to 547mg per mL
A280: 0.04 to 750mg per mL
Lysozyme: 0.015 to 284mg per mL |
| Detector Type |
2048-element linear CCD array |
| Sample Stage Material |
Quartz fiber and 304 stainless steel |
| Input Voltage |
12Vdc, 4A |
| Power Consumption |
48W |
| Dimensions |
270x210x196mm |
| Weight |
3.8kg |
| OD600 Measurements |
| Absorbance Range |
0 to 4.000Abs |
| Absorbance Stability |
(0, 3): ≤ 0.5%, [3, 4): ≤ 1.5% |
| Absorbance Repeatability |
(0, 3): ≤ 0.5%, [3, 4): ≤ 1.5% |
| Absorbance Accuracy |
(0, 2): ≤ 0.005A, [2, 3): ≤ 1%, [3, 4): ≤ 2% |
| Heating Temperature |
37°C |
| Stirring Speed |
9 levels, 100 to 900rpm |
| Fluorescence Detection |
| Sample Volume |
1 to 20uL |
| Detection Time |
3 seconds |
| Detection Range |
0.01 to 120ng per uL dsDNA HS, 0.2 to 2000ng per uL dsDNA BR, 0.05 to 240ng per uL Oligo, 0.1 to 20mg
per mL Protein BR |
| Repeatability |
less than 1.5% |
| Dynamic Range |
5 orders of magnitude |
| Linearity |
Coefficient of Determination ≥ 0.995 |
| Light Source |
Monochromatic LED |
| Excitation Wavelengths |
470nm, 625nm, standard
365nm, 525nm, optional |
| Emission Wavelengths |
525nm, 690nm, standard
460nm, 620nm, optional |